Introduction:

Hematology analyzer is a medical device used to detect and analyze the cellular components of blood, mainly including white blood cells, red blood cells, platelets and so on. It uses optical, electronic and other technologies to count, classify and observe the morphology of blood samples. Blood cell analyzer has important application value in clinical examination, disease prevention, diagnosis and treatment.

Quantity:

Model

Features:

  1. 9.7-inch color touch LCD screen, easy to operate, display all parameters and histograms on the same screen.

  2. Fewer samples are required for testing: venous blood: 10μl, capillary: 10ul, prediluted: 20μl.

  3. 60 samples/hour.

  4. 21 parameters+3 histograms.

  5. It can be turned on continuously for 24 hours.

  6. Automatically clean the sampler, counting pool and pipeline after starting up.

  7. Built-in thermal printer, external printer can be connected.

  8. Electrical impedance method, cyanide-free detection of HGB.

  9. Fully intelligent blockage removal and alarm function.

  10. Both automatic calibration and manual calibration are available.

  11. Counting Mode: Venous, capilary, prediluted.

  12. Support LIS system.

Technical Parameters:

Model


BH-HA310

Throughput

60 samples/hour

Principle

Impedance for WBC differentiation and WBC/RBC/PLT count; Colorimetric method for HGB

Channels

2

Parameters

21 parameters(including WBC, Mid#, Lym#, Gran#, Mid%, Lym%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR, P-LCC)

Histograms

3 histograms(including WBC, RBC, PLT histograms)

Sample Volume

Venous blood: 10μl, Capillary: 10μl, Prediluted: 20μl

Calibration

Manual, auto calibration

Counting Modes

Venous, capilary, prediluted

Printing Model

Auto print, manual print

LIS

Support Bi-directional LIS

Language

Chinese, English, Spanish, French, Russian, Ukrainian

Reagent

Diluent 10L, Hemolytic agent 250ml, Probe cleanser: 50ml

Data Input

9.7-inch touch screen, mouse, keyboard(optional)

Output

Internal printer, support external printer

Printer Paper

57*30mm

Interface

4 USB ports, network port, RS232 serial port

Storage

100,000 results with histogram

Unclog

High voltage burn, reverse high-pressure flush

Work Temperature

10~40°C

Power Supply

100~240AC, 50/60Hz, 150VA

Relative Humidity

≤80%

Atmospheric Pressure

86.0kPa~106.0kPa

Carry-over Rate

Parameter

WBC

RBC

HGB

PLT

CV

≤1.5%

≤1%

≤1%

≤3%

Background

Parameter

WBC

RBC

HGB

HCT

PLT

Background

≤0.5*10^9/L

≤0.05*10^12/L

≤2g/L

≤0.5%

≤10*10^9/L

Linearity

Parameter

Measurement Range

CV

WBC

(1~10.00*10^9)/L

(10.10~99.9*10^9)/L

(±0.5*10^9)/L

±5%

RBC

(0.3~1.00*10^12)/L (1.01~7*10^12)/L

(±0.05*10^12)/L

±5%

HGB

(20~70g)/L

(71~200.0g)/L

(±2.0g)/L

±3%

PLT

(20~100*10^9)/L

(101~999*10^9)/L

(±10.0*10^9)/L

±10%

Precision

Parameter

Repeatability

Measurement Range

WBC

≤4.0%

(3.5~9.5*10^9)/L

RBC

≤2.0%

(3.8~5.8*10^12)/L

HGB

≤2.0%

(115~175)g/L

PLT

≤8.0%

(125~350*10^9)/L

HCT

≤3.0%

35%~50%

External Size(W*D*H)

295*505*430mm

Packed Size(W*D*H)

Instrument: 404*564*616mm; Reagent: 340*340*340mm, 200*150*100mm

Net Weight

20kg

Gross Weight

Instrument: 23kg; Reagent: 13kg, 1kg